Preventative Maintenance Kits G5010007) and spare part kits — Class II medical device recall Z-1280-2021
Terminated recall by Baxter Healthcare Corporation, reported 2021-03-31, distributed nationwide. Variability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning Syst
| Recall number | Z-1280-2021 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Baxter Healthcare Corporation, Deerfield, IL, United States |
| Recall initiated | 2021-01-28 |
| Classified | 2021-03-19 |
| Reported by FDA | 2021-03-31 |
| Terminated | 2024-06-13 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AR, AT, AU, BB, BE, BG, BH, BM, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EE, EG, ES, FI, FR, GB, GR, GT, HK, HR, HT, HU, IE, IL, IN, IQ, IR, IS, IT, JO, JP, KE, KR, KW, KY, LB, LT, LV, MA, MT, MX, MY, NG, NL, NO, NZ, OM, PK, PL, PT, QA, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TT, TW, UA, VE, VN, ZA |
| Quantity | 35,175 kits |
Product
Preventative Maintenance Kits G5010007) and spare part kits (G5064801 and G5006203) - Product Usage: For Prismaflex Control Units, used for Continuous Renal Replacement Therapy (CRRT).
Reason for recall
Variability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning System) Pump Assembly, may lead to alarm situations during or after a system self-test.