Raystation treatment planning system for radiation therapy — Class II medical device recall Z-1281-2023
Ongoing recall by RAYSEARCH LABORATORIES AB, reported 2023-03-29, distributed nationwide. It is possible to set a non-zero collimator angle in the Virtual Simulation module. The angle will be correct
| Recall number | Z-1281-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | RAYSEARCH LABORATORIES AB, Stockholm, N/A, Sweden |
| Recall initiated | 2023-02-17 |
| Classified | 2023-03-22 |
| Reported by FDA | 2023-03-29 |
| Distributed | Nationwide (US) |
| Quantity | 10,770 systems |
| Model numbers | 10.0.0, 10.0.1, 10.0.2, 10.1.0, 10.1.1, 11.0.0, 11.0.1, 11.0.3, 11.0.4, 12.0.0, 12.0.3, 12.0.4, 12.1.0, 12.1.1, 12.1.2, 12.1.3, 13.0.0, 13.1.0, 9.0.0, 9.0.1, 9.1.0, 9.2.0 |
Product
Raystation treatment planning system for radiation therapy, Model Numbers 9.0.0, 9.0.1, 9.1.0, 9.2.0, 10.0.0, 10.0.1, 10.0.2, 10.1.0, 10.1.1, 11.0.0, 11.0.1, 11.0.3, 11.0.4, 12.0.0, 12.1.0, 12.1.1, 12.0.3, 12.1.2, 12.0.4, 12.1.3, 13.0.0, 13.1.0. Software Version Numbers 9A, 9B, 10A, 10B, 11A, 11B and 12A including service packs
Reason for recall
It is possible to set a non-zero collimator angle in the Virtual Simulation module. The angle will be correct in the user interface but if the plan is DICOM exported by using the VSIM Export button, the angle will not be correct in the exported RT plan. In the exported RT plan, the DICOM attribute Beam Limiting Device Angle (300A,0120) is set to zero, regardless of the selected collimator angle.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1281-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.