Data Foundry

Medical device recalls 2019

Guardian Connect App CSS7200 used on the iPhone — Class II medical device recall Z-1282-2019

Terminated recall by Medtronic Inc., reported 2019-05-15. The application may be closed by the operating system without alerting the user the app is no longer running o

Recall numberZ-1282-2019
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMedtronic Inc., Northridge, CA, United States
Recall initiated2018-07-04
Classified2019-05-03
Reported by FDA2019-05-15
Terminated2022-06-06
Countries outside the USAT, AU, BE, CH, CL, CO, CZ, DE, DK, ES, FI, FR, GB, HU, IE, IL, IS, IT, KR, LU, NL, NO, PL, QA, RO, SA, SE, SG, SI, SK, TR, ZA
Quantity9,328 registered
UPC / GTIN / UDI-DI00643169693142

Product

Guardian Connect App CSS7200 used on the iPhone, iPad, and iPod Touch devices.

Reason for recall

The application may be closed by the operating system without alerting the user the app is no longer running or communicating with the transmitter resulting in the user not receiving alerts that could be associated with hypoglycemic or hyperglycemic events.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1282-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.