Data Foundry

Medical device recalls 2022

QMS Tacrolimus Immunoassay REF 10015573 CALIBRATOR SET QMS TACROLIMUS… — Class II medical device recall Z-1284-2022

Ongoing recall by Microgenics Corporation, reported 2022-07-06, distributed in AL, AZ, CA, DC, FL, IA, ID, IL…. Due to QMS Tacrolimus Calibrator recovering between 82 to 94% of their target values, resulting in over-quanti

Recall numberZ-1284-2022
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMicrogenics Corporation, Fremont, CA, United States
Recall initiated2022-01-20
Classified2022-06-24
Reported by FDA2022-07-06
DistributedAL, AZ, CA, DC, FL, IA, ID, IL, KY, LA, MA, NE, NJ, NM, OH, OK, PA, SC, TN, TX, VA
Countries outside the USBD, BR, CA, DE, IN, KR, MX, MY, TH
Quantity615 kits
Lot numbers74175462, 74267412
Model numbers10015573

Product

QMS Tacrolimus Immunoassay REF 10015573 CALIBRATOR SET QMS TACROLIMUS CAL A 10015566-3 QMS TACROLIMUS CAL B 10015567-3 QMS TACROLIMUS CAL C 10015568-3 QMS TACROLIMUS CAL D 10015569-3 QMS TACROLIMUS CAL E 10015570-3 QMS TACROLIMUS CAL F 10015571-3

Reason for recall

Due to QMS Tacrolimus Calibrator recovering between 82 to 94% of their target values, resulting in over-quantitation in tacrolimus patient result values when used with the QMS Tacrolimus Immunoassay.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1284-2022. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.