Biosense Webster — Class II medical device recall Z-1286-2021

Completed recall by Biosense Webster, Inc., reported 2021-03-31. There is a potential for the hemostatic valve to dislodge while introducing the dilator or device into the she

Recall numberZ-1286-2021
Product typeMedical device
ClassificationClass II
StatusCompleted
Voluntary or mandatedVoluntary
Recalling firmBiosense Webster, Inc., Irwindale, CA, United States
Recall initiated2021-02-26
Classified2021-03-22
Reported by FDA2021-03-31
UPC / GTIN / UDI-DI10846835016260
Model numbersD138503

Product

Biosense Webster, CARTO VIZIGO, Bi-Directional Guiding Sheath, 8.5F (Large) C3, REF: D138503, Rx Only, CE 2797, STERILE EO, udi: (01)10846835016260 - Product Usage: provide accessibility and maneuverability in the cardiac anatomy during cardiac electrophysiological mapping and ablation procedures.

Reason for recall

There is a potential for the hemostatic valve to dislodge while introducing the dilator or device into the sheath. This could result in minor bleeding and/or embolism