Uterine Electromyographic Monitor -Monica IF24 Interface System… — Class II medical device recall Z-1293-2018
Terminated recall by GE Healthcare, LLC, reported 2018-04-11, distributed nationwide. if the cables of the Monica IF24 System are removed by a user, and then incorrectly re-connected the Fetal Hea
| Recall number | Z-1293-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | GE Healthcare, LLC, Waukesha, WI, United States |
| Recall initiated | 2017-12-15 |
| Classified | 2018-04-02 |
| Reported by FDA | 2018-04-11 |
| Terminated | 2020-04-21 |
| Distributed | Nationwide (US) |
| Quantity | 355 |
| Model numbers | 105-PT-001 |
Product
Uterine Electromyographic Monitor -Monica IF24 Interface System Product Usage: The Monica IF24 CTG Interface Device is an accessory to the Monica AN24 which provides a means of interfacing the wireless output of the Monica AN24 to the transducer inputs of a CTG Fetal monitor. The Monica IF24 CTG Interface Device enables signals collected by the Monica AN24 to be printed and displayed on a CTG Fetal Monitor and sent on to a central network, if connected. The device is intended for use by healthcare professionals in a clinical setting.
Reason for recall
if the cables of the Monica IF24 System are removed by a user, and then incorrectly re-connected the Fetal Heart Rate (FHR) and Maternal Heart Rate (MHR) signals would be displayed incorrectly (swapped).