Data Foundry

Medical device recalls 2018

Uterine Electromyographic Monitor -Monica IF24 Interface System… — Class II medical device recall Z-1293-2018

Terminated recall by GE Healthcare, LLC, reported 2018-04-11, distributed nationwide. if the cables of the Monica IF24 System are removed by a user, and then incorrectly re-connected the Fetal Hea

Recall numberZ-1293-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmGE Healthcare, LLC, Waukesha, WI, United States
Recall initiated2017-12-15
Classified2018-04-02
Reported by FDA2018-04-11
Terminated2020-04-21
DistributedNationwide (US)
Quantity355
Model numbers105-PT-001

Product

Uterine Electromyographic Monitor -Monica IF24 Interface System Product Usage: The Monica IF24 CTG Interface Device is an accessory to the Monica AN24 which provides a means of interfacing the wireless output of the Monica AN24 to the transducer inputs of a CTG Fetal monitor. The Monica IF24 CTG Interface Device enables signals collected by the Monica AN24 to be printed and displayed on a CTG Fetal Monitor and sent on to a central network, if connected. The device is intended for use by healthcare professionals in a clinical setting.

Reason for recall

if the cables of the Monica IF24 System are removed by a user, and then incorrectly re-connected the Fetal Heart Rate (FHR) and Maternal Heart Rate (MHR) signals would be displayed incorrectly (swapped).

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1293-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.