Data Foundry

Medical device recalls 2019

Medica Wash1-Wash solution used prior to Lipase assay on the EasyRA… — Class II medical device recall Z-1294-2019

Terminated recall by Medica Corporation, reported 2019-05-15, distributed nationwide. Wash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Li

Recall numberZ-1294-2019
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMedica Corporation, Bedford, MA, United States
Recall initiated2019-02-06
Classified2019-05-07
Reported by FDA2019-05-15
Terminated2020-10-02
DistributedNationwide (US)
Countries outside the USEE, FR, GT, IN, IQ, IR, KE
Quantity87 kits
Reason classespotency
Lot numbers18066, 18133, 18235, 18317
Model numbers10680

Product

Medica Wash1-Wash solution used prior to Lipase assay on the EasyRA clinical chemistry analyzer REF 10680 - Product Usage: For use in washing the probe of the Medica EasyRA Chemistry Analyzer. The wash solution in the wedge is used to prevent interference between certain tests.

Reason for recall

Wash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Lipase (LIP) assay on the EasyRA clinical chemistry analyzer.