Data Foundry

Medical device recalls 2025

TISSUE APPROXIMATION SYSTEM — Class II medical device recall Z-1294-2025

Ongoing recall by TAS Medical Inc, reported 2025-03-12, distributed nationwide. Tissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim

Recall numberZ-1294-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmTAS Medical Inc, Windermere, FL, United States
Recall initiated2025-01-27
Classified2025-03-06
Reported by FDA2025-03-12
DistributedNationwide (US)
Quantity114
Reason classessterility
UPC / GTIN / UDI-DI00197644217267, 00197644927241, 00860011435914, 00860011435921
Lot numbersT-LAP/00860011435914
Model numbersT-4000, T-5000

Product

TISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP

Reason for recall

Tissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties in 104-degrees F sterile saline solution for 60 min prior to implantation to rehydrate the zip-tie straps. Broken zip-ties not observed during procedure could lead to future hernia recurrence, recurrent surgery to correct a large herniation.