Cios Alpha with software version VA30-mobile C-arm x-ray Material #… — Class II medical device recall Z-1295-2019

Terminated recall by Siemens Medical Solutions USA, Inc, reported 2019-05-15, distributed nationwide. Continuous operation of the Cios Alpha at high tube outputs may result in increased wear of radiation emitting

Recall numberZ-1295-2019
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Medical Solutions USA, Inc, Malvern, PA, United States
Recall initiated2019-03-22
Classified2019-05-07
Reported by FDA2019-05-15
Terminated2020-05-12
DistributedNationwide (US)
Quantity2 us

Product

Cios Alpha with software version VA30-mobile C-arm x-ray Material # 11105200 Product Usage: The Cios Alpha is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients

Reason for recall

Continuous operation of the Cios Alpha at high tube outputs may result in increased wear of radiation emitting components, risk of premature failure of the Monoblock and loss of imaging X-rays exists

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1295-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.