VITROS Chemistry Products Calibrator Kit 4- — Class II medical device recall Z-1295-2021
Terminated recall by Ortho-Clinical Diagnostics, Inc., reported 2021-03-31, distributed nationwide. Potential total iron-binding capacity (TIBC) results. The VITROS TIBC performance shift will start with the ne
| Recall number | Z-1295-2021 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ortho-Clinical Diagnostics, Inc., Rochester, NY, United States |
| Recall initiated | 2021-02-18 |
| Classified | 2021-03-23 |
| Reported by FDA | 2021-03-31 |
| Terminated | 2022-10-18 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, BE, BR, CA, CL, CN, CO, DE, DK, ES, FR, GB, IN, IT, JP, MX, NL, NO, PL, PT, RU, SE, SG |
| UPC / GTIN / UDI-DI | 10758750000289, 10758750002276, 10758750008537, 10758750009244 |
| Lot numbers | 0449 |
| Expiration dates | 2021-08-28, 2022-02-10, 2022-10-01 |
Product
VITROS Chemistry Products Calibrator Kit 4-(Calibrator values for TIBC only). Used to calibrate VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems Product Code: 1204668 Impacted Products: A. VITROS Chemistry Products TIBC Kit Product No: 7541 (UDI 8750005178) All current and new lots B. VITROS Chemistry Products Fe Slides Product No: 1924547 90 Slides (UDI 10758750000289) Product No: 1515808 300 slides (UDI 10758750009244) Product No: 6802198 (Japan) 90 Slides (UDI 10758750002276) All current, new and previous lots
Reason for recall
Potential total iron-binding capacity (TIBC) results. The VITROS TIBC performance shift will start with the newest Generation (GEN) of VITROS Fe Slides (GEN 26), the potential shift in TIBC patient values of GEN 26 and above will be approximately 80 ug/dL lower across the measuring range than other available GENS of VITROS Fe Slides (GENs 22-25).
Machine access