Operating Table System — Class II medical device recall Z-1296-2021

Ongoing recall by Getinge Group Logistics America, LLC, reported 2021-03-31, distributed in AK, AL, AR, AZ, CA, CO, DE, FL…. There is a potential for hydraulic leakage at the inclination function to occur resulting in an unintended Tre

Recall numberZ-1296-2021
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmGetinge Group Logistics America, LLC, Wayne, NJ, United States
Recall initiated2021-02-11
Classified2021-03-23
Reported by FDA2021-03-31
DistributedAK, AL, AR, AZ, CA, CO, DE, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV
Countries outside the USAE, AR, AT, AU, BE, BG, BH, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GB, GH, GR, HK, HR, HU, ID, IE, IL, IN, IT, JO, JP, KR, KW, LB, LT, LU, LV, MA, MX, MY, NL, NO, NZ, OM, PA, PH, PL, PT, QA, RO, RS, RU, SA, SE, SG, SI, TH, TR, TW, VE, VN, ZA
Quantity4,288 units
Reason classespackaging
UPC / GTIN / UDI-DI04046768031500, 04046768031531, 04046768031555, 04046768031579, 04046768040700, 04046768088061, 04046768088351, 04046768124752
Serial numbers1028, 1159, 1264, 3402, 401, 762
Model numbers118001A0, 118001B0, 118001B1, 118001B2, 118001B3, 118001B4, 118001C0, 118001D0

Product

Operating Table System

Reason for recall

There is a potential for hydraulic leakage at the inclination function to occur resulting in an unintended Trendelenburg or anti-Trendelenburg movement of the table top depending on patient positioning which may lead to the patient having to be transferred to another OR table causing a delay in the procedure.