Data Foundry

Medical device recalls 2020

Guided Start Drills/Counterbore NobRpl NP — Class II medical device recall Z-1302-2020

Terminated recall by Nobel Biocare Usa Llc, reported 2020-02-26, distributed nationwide. Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaran

Recall numberZ-1302-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmNobel Biocare Usa Llc, Yorba Linda, CA, United States
Recall initiated2019-03-14
Classified2020-02-19
Reported by FDA2020-02-26
Terminated2024-02-15
DistributedNationwide (US)
Reason classessterility
UPC / GTIN / UDI-DI07332747005295
Model numbers32737

Product

Guided Start Drills/Counterbore NobRpl NP - Product Usage: The Guided Drills (Twist (Step) Drills, Guided Tapered Drills) are used for preparing osteotomies for the specific related implants

Reason for recall

Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.