Data Foundry

Medical device recalls 2026

Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog… — Class II medical device recall Z-1304-2026

Ongoing recall by Fujirebio Diagnostics, Inc., reported 2026-02-11, distributed nationwide. Customers may experience inaccurate test results by observing a higher frequency of classification in the inde

Recall numberZ-1304-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmFujirebio Diagnostics, Inc., Malvern, PA, United States
Recall initiated2025-12-11
Classified2026-02-05
Reported by FDA2026-02-11
DistributedNationwide (US)
Quantity148 units
UPC / GTIN / UDI-DI05414470815586
Lot numbersD6C6092U, D6C6092U1, D6C6123U
Model numbers81558

Product

Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Product Description: Lumipulse pTau 217 Plasma Controls: This quality control product can be used as an objective judgement of the laboratory s procedures and personnel techniques when analyzing pTau 217 in K2EDTA plasma on the LUMIPULSE G System. It is a valuable tool to assess good laboratory practices. Two levels of control are available to compare observations with expected ranges therefore assuring consistent performance of the testing system within the clinical range. Lumipulse pTau 217 Plasma Controls: Liquid (Frozen), 2 ¿2 Concentrations L1 Level 1 (2¿ 1.5 mL) L2 Level 2 (2¿ 1.5 mL) Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 300, 0.05% ProClin 950 Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio

Reason for recall

Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1304-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.