Data Foundry

Medical device recalls 2026

Product Name: Lumipulse ¿-Amyloid Plasma Controls Model/Catalog… — Class II medical device recall Z-1307-2026

Ongoing recall by Fujirebio Diagnostics, Inc., reported 2026-02-11, distributed nationwide. Customers may experience inaccurate test results by observing a higher frequency of classification in the inde

Recall numberZ-1307-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmFujirebio Diagnostics, Inc., Malvern, PA, United States
Recall initiated2025-12-11
Classified2026-02-05
Reported by FDA2026-02-11
DistributedNationwide (US)
Quantity179 units
UPC / GTIN / UDI-DI05414470815593
Lot numbersT8B6116U
Model numbers81559

Product

Product Name: Lumipulse ¿-Amyloid Plasma Controls Model/Catalog Number: 81559 Software Version: Not Applicable Product Description: Lumipulse ¿-Amyloid Plasma Controls: This quality control product can be used as an objective judgement of the laboratory s procedures and personnel techniques when analyzing ¿-amyloid1-42 in K2EDTA plasma on the LUMIPULSE G System. It is a valuable tool to assess good laboratory practices. Two levels of control are available to compare observations with expected ranges therefore assuring consistent performance of the testing system within the clinical range. Lumipulse ¿-Amyloid Plasma Controls: Liquid (Frozen), 2 ¿2 Concentrations L1 Level 1 (2¿ 1.5 mL) L2 Level 2 (2¿ 1.5 mL) Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 950 Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio

Reason for recall

Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).