Data Foundry

Medical device recalls 2016

VERO MHI-TM2000 Linear Accelerator System --- Common Name: Medical… — Class II medical device recall Z-1308-2016

Terminated recall by MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK, reported 2016-04-06, distributed nationwide. In rare cases the VERO/MHI-TM2000 Operator Console could set an incorrect Gantry angle for the first beam of a

Recall numberZ-1308-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK, HIROSHIMA, N/A, Japan
Recall initiated2016-03-07
Classified2016-03-31
Reported by FDA2016-04-06
Terminated2017-04-20
DistributedNationwide (US)
Serial numbers201902, 203901, 203919, 203924

Product

VERO MHI-TM2000 Linear Accelerator System --- Common Name: Medical Linear Accelerator MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.

Reason for recall

In rare cases the VERO/MHI-TM2000 Operator Console could set an incorrect Gantry angle for the first beam of an IMRT treatment if certain specific conditions are met. If Gantry angle for the first beam of IMRT is incorrectly set and if nobody is aware of it, an erroneous treatment would be given to the patient.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1308-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.