Data Foundry

Medical device recalls 2024

NautilusTM Smart ECMO Module with Balance Biosurface-used in… — Class II medical device recall Z-1310-2024

Ongoing recall by Mc3 Inc, reported 2024-03-27, distributed nationwide. May have increased water path restriction within the heat exchanger, leading to degraded heat exchange perform

Recall numberZ-1310-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMc3 Inc, Dexter, MI, United States
Recall initiated2023-10-11
Classified2024-03-19
Reported by FDA2024-03-27
DistributedNationwide (US)
Quantity188 units
UPC / GTIN / UDI-DI10854916006871, 10854916006895

Product

NautilusTM Smart ECMO Module with Balance Biosurface-used in extracorporeal life support procedures to oxygenate blood, remove carbon dioxide and regulate blood temperature. Product Number: 48135, 48135E

Reason for recall

May have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange performance may be experienced immediately upon priming or may worsen over the duration of use. The potential harm as stated in the IFU includes hypothermia due to inadequate heat transfer