Reprocessed Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound… — Class I medical device recall Z-1324-2026
Ongoing recall by Medline Industries, LP, reported 2026-02-18, distributed nationwide. These lots of reprocessed catheters may contain small particles of residual material. If affected devices are
| Recall number | Z-1324-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medline Industries, LP, Northfield, IL, United States |
| Recall initiated | 2025-12-22 |
| Classified | 2026-02-06 |
| Reported by FDA | 2026-02-18 |
| Distributed | Nationwide (US) |
| Quantity | 94 |
| UPC / GTIN / UDI-DI | 10197344157273, 10197344157280, 10197344157297, 10197344157303, 20197344157287 |
| Lot numbers | 10438577RH, 10439011RH, 10439236RH, EP250724, EP250808 |
Product
Reprocessed Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheters: Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheter Use on GE Systems 10439072RH; Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheter Use on Siemens Systems 10438577RH; Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheter, Use on Siemens Systems 10439011RH
Reason for recall
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025