Data Foundry

Medical device recalls 2021

Philips SureSigns VM4 — Class II medical device recall Z-1326-2021

Terminated recall by Philips North America, LLC, reported 2021-04-07. Fail to Comply with Chinese Standard YY1079-2008: Clauses 4.2.6 and 4.2.7.3. Range/accuracy of heart rate mete

Recall numberZ-1326-2021
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips North America, LLC, Andover, MA, United States
Recall initiated2021-03-23
Classified2021-03-30
Reported by FDA2021-04-07
Terminated2023-09-28
Countries outside the USCN
Quantity31,773 units
UPC / GTIN / UDI-DI00453564024461
Model numbers453564021961, 453564021971, 453564022041, 453564024371, 453564024401, 453564024421, 453564024441, 453564024461, 453564024481, 53564022001, 53564022081, 863063, 863064, 863065, 863066, 863068, 863085, 863086

Product

Philips SureSigns VM4, VM6 and VM8 - Product Usage: patient monitors-for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. (CHINA only) SureSigns VM4 (863063, 863085) SureSigns VM6 (863064, 863065, 863086) SureSi"gns VM8 (863066, 863068, 863087, 863088) Model Numbers: 863063,863064,863065,863066,863068,863085,863086, 453564021961,453564021971, 53564022001,453564022041, 53564022081,453564024371, 453564024401,453564024421 453564024441,453564024461, 453564024481

Reason for recall

Fail to Comply with Chinese Standard YY1079-2008: Clauses 4.2.6 and 4.2.7.3. Range/accuracy of heart rate meter for pediatric mode-In pediatric mode, when the input signal rate is over 300 bpm, the indicated rate of the affected products will be lower than this upper limit