Data Foundry

Medical device recalls 2014

Spirit MB Bracket — Class II medical device recall Z-1330-2014

Terminated recall by Ormco/Sybronendo, reported 2014-04-09. Ormco Corporation initiated this recall of Spirit MB Brackets, Part Number 494-0110, Lot Number 081351013, bec

Recall numberZ-1330-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOrmco/Sybronendo, Glendora, CA, United States
Recall initiated2014-03-25
Classified2014-04-03
Reported by FDA2014-04-09
Terminated2014-04-18
Countries outside the USJP
Quantity150 units
Lot numbers081351013
Model numbers494-0110

Product

Spirit MB Bracket, UR Central +14T +5A.018, Part Number 494-0110, Lot Number 081351013. Product Usage: An orthodontic bracket. Spirit MB Brackets are a fixed attachment on a tooth which holds an archwire during orthodontic treatment.

Reason for recall

Ormco Corporation initiated this recall of Spirit MB Brackets, Part Number 494-0110, Lot Number 081351013, because the ID dot, which designates the bracket as distal/gingival, is misplaced. The misidentification may lead an orthodontist to place the bracket on the wrong tooth, leading to movement 10 degrees in the wrong direction.