Data Foundry

Medical device recalls 2021

Product Label — Class II medical device recall Z-1330-2021

Ongoing recall by Siemens Medical Solutions USA, Inc., reported 2021-04-07, distributed in AK, AL, AR, AZ, CA, CT, DE, FL…. The ultrasound system averages the Mean Sac Diameter (MSD) and Gestational Sac Diameter (GSD) into the Estimat

Recall numberZ-1330-2021
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmSiemens Medical Solutions USA, Inc., Issaquah, WA, United States
Recall initiated2021-02-04
Classified2021-04-01
Reported by FDA2021-04-07
DistributedAK, AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY
Countries outside the USAE, AT, AU, BE, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DZ, ES, FI, FR, GB, GR, HK, HR, ID, IL, IN, IR, IT, JP, KR, KW, MA, MX, NL, NZ, OM, PH, PL, PT, QA, RO, SA, SE, SG, SI, SK, TH, TR, TW, UA, VN, ZA
Quantity1,172 systems
Serial numbers800015, 800101, 800103, 800104, 800105, 800106, 800107, 800108, 800109, 800111, 800112, 800113, 800114, 800115, 800116, 800117, 800118, 800119, 800120, 800121, 800122, 800123, 800126, 800127, 800129 and 475 more
Model numbers11148775

Product

Product Label - ACUSON Sequoia *** Power Input: 100-240V~, 13.0-5.4A, 50-60Hz *** (240)11148775 Model User Manual states - ACUSON Sequoia Diagnostic Ultrasound Systems

Reason for recall

The ultrasound system averages the Mean Sac Diameter (MSD) and Gestational Sac Diameter (GSD) into the Estimated Date of Delivery (EDD) calculation. This may result in an incorrect EDD, which may influence patient management decisions regarding induction of labor and elective caesarean delivery, which may result in premature births.