Terminated recall by Siemens Medical Solutions USA, Inc., reported 2014-04-09. ACUSON P300 ultrasound systems using the PA230 transducer with a user-defined preset which includes the 100% s

Recall numberZ-1331-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Medical Solutions USA, Inc., Mountain View, CA, United States
Recall initiated2014-03-07
Classified2014-04-03
Reported by FDA2014-04-09
Terminated2015-11-12
Countries outside the USAE, AT, AU, BE, BG, BO, CA, CH, CL, CO, DE, DK, EC, ES, FR, GB, GR, HR, ID, IE, IL, IN, IT, JP, KR, KW, MA, MY, NL, NO, PE, PH, PL, PT, RO, RS, SA, SG, TR, TT, TW, ZA
Quantity675
Serial numbers105, 106, 114, 116, 117, 118, 119, 120, 121, 122, 123, 124, 125, 126, 127, 128, 129, 130, 131, 132, 133, 134, 135, 136, 137 and 475 more
Model numbers10433816, 10852384, 320E, P300PA

Product

Siemens Medical Solutions, ACUSON P300 Ultrasound system with P300 PA 230E cardiac probe (Esaote product 7348). Provides imaging for guidance of biopsy and imaging to assist in the placement of needles in vascular or other anatomical structures as well as peripheral nerve blocks in Musculoskeletal applications.

Reason for recall

ACUSON P300 ultrasound systems using the PA230 transducer with a user-defined preset which includes the 100% setting for output power could cause the transducer surface temperature to exceed the maximum allowable limit.