Data Foundry

Medical device recalls 2020

Luminos Agile Max — Class II medical device recall Z-1332-2020

Terminated recall by Siemens Medical Solutions USA, Inc, reported 2020-02-26, distributed nationwide. Siemens became aware of an issue of incorrect use of the bypass key overriding the built-in collision control

Recall numberZ-1332-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Medical Solutions USA, Inc, Malvern, PA, United States
Recall initiated2020-01-08
Classified2020-02-20
Reported by FDA2020-02-26
Terminated2022-06-21
DistributedNationwide (US)
Quantity475
Serial numbers61013, 61014, 61015, 61017, 61018, 61021, 61022, 61025, 61027, 61028, 61029, 61030, 61031, 61032, 61033, 61034, 61035, 61036, 61037, 61038, 61039, 61040, 61041, 61042, 61043 and 450 more
Model numbers10762472

Product

Luminos Agile Max, Model No. 10762472 - Product Usage: Luminos Agile Max is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image. The system has medical applications ranging from gastrointestinal examinations to cranial, skeletal, thoracic and lung exposures as well as examinations of the urogenital tract.

Reason for recall

Siemens became aware of an issue of incorrect use of the bypass key overriding the built-in collision control creating risk of collision with obstacles or persons when the operator uses the override function.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1332-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.