Dexcom G4 PLATINUM — Class I medical device recall Z-1335-2016
Terminated recall by Dexcom Inc, reported 2016-04-20, distributed nationwide. Customers may not receive an intended audible alarm or alert if relying on hearing the alarm or alert.
| Recall number | Z-1335-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Dexcom Inc, San Diego, CA, United States |
| Recall initiated | 2016-02-23 |
| Classified | 2016-04-11 |
| Reported by FDA | 2016-04-20 |
| Terminated | 2017-08-11 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AT, AU, BE, BH, CA, CH, CL, CO, CZ, DE, DK, ES, FI, FR, GB, HK, HU, IE, IL, IN, IT, KW, LB, NL, NO, NZ, PL, PT, QA, SE, SI, TR, ZA |
| Quantity | 18,004 |
| Lot numbers | 5120293, 5120294, 5120309, 5120310, 5120311, 5120312, 5126237, 5126238, 5126239, 5135361, 5135362, 5135363, 5135364, 5135365, 5135366, 5135367, 5135368, 5135369, 5135370, 5135371, 5135379, 5135380, 5135381, 5136498, 5136500, 5136502, 5136504, 5136505, 5139379, 5139380, 5139381, 5139382, 5139386, 5142406, 5142408, 5142415, 5142416, 5142417, 5150416, 5150417 and 183 more |
| Model numbers | MT22430 |
Product
Dexcom G4 PLATINUM (Pediatric) Receiver Receiver Part Number: MT22430 Receiver (mg/dL), Black (pediatrics); MT22430-PNK, Receiver (mg/dL), Pink (pediatrics); MT22430-BLU, Receiver (mg/dL); Blue (pediatrics). Intended to detect trends and track glucose patterns in persons with diabetes.
Reason for recall
Customers may not receive an intended audible alarm or alert if relying on hearing the alarm or alert.