Data Foundry

Medical device recalls 2024

MINI STICK MAX 5F X 10 CM STD — Class II medical device recall Z-1339-2024

Ongoing recall by Angiodynamics, Inc., reported 2024-03-27, distributed nationwide. Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure.

Recall numberZ-1339-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmAngiodynamics, Inc., Queensbury, NY, United States
Recall initiated2024-01-30
Classified2024-03-21
Reported by FDA2024-03-27
DistributedNationwide (US)
Quantity5,176 units
UPC / GTIN / UDI-DI15051684022996
Lot numbers5796084, 5796085, 5796086, 5796087, 5796088, 5796089, 5796090, 5796659, 5796660, 5796661, 5796662, 5796663, 5796666, 5797456, 5797465, 5798836, 5798841, 5798842, 5799620, 5800404, 5803526, 5804396
Model numbers45-756

Product

MINI STICK MAX 5F X 10 CM STD .018 NI/TU ECHO 2.75" PG Catalog Number: 45-756

Reason for recall

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1339-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.