NAMIC CONVENIENCE KIT RENOWN REGIONAL MEDICAL CENTER — Class II medical device recall Z-1344-2016
Terminated recall by AngioDynamics Inc., reported 2016-04-13, distributed nationwide. Procedure Products initiated the recall due to the inner lumen of the needles not meeting the intended specifi
| Recall number | Z-1344-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | AngioDynamics Inc., Glens Falls, NY, United States |
| Recall initiated | 2015-11-03 |
| Classified | 2016-04-07 |
| Reported by FDA | 2016-04-13 |
| Terminated | 2017-03-16 |
| Distributed | Nationwide (US) |
| Quantity | 100 kits |
| UPC / GTIN / UDI-DI | 00965652300291 |
| Lot numbers | 2018-08, 4925152, 4930012 |
| Model numbers | 65230029 |
Product
NAMIC CONVENIENCE KIT RENOWN REGIONAL MEDICAL CENTER, REF Catalog No. 65230029, UPN Product No. H965652300290, Product is packed as a box of 30 pieces, box label contains UPN Product No. H965652300291 Product Usage - This needle is intended to be placed over the guidewire to enlarge the opening in the vessel, and is then removed before sliding the catheter over the guidewire
Reason for recall
Procedure Products initiated the recall due to the inner lumen of the needles not meeting the intended specification. Specifically, at least a portion of the affected lot is unable to accept a 0.038" guidewire as intended. In some cases the inner lumen was found to be 0.036".