THP Hip Plating System — Class II medical device recall Z-1347-2020
Terminated recall by Zimmer Biomet, Inc., reported 2020-03-04, distributed in AL, CA, FL, IA, IL, MD, MI, MN…. Potential increase in fracture rate was detected during internal testing, in addition to potential discolorati
| Recall number | Z-1347-2020 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Zimmer Biomet, Inc., Warsaw, IN, United States |
| Recall initiated | 2020-01-07 |
| Classified | 2020-02-24 |
| Reported by FDA | 2020-03-04 |
| Terminated | 2021-04-08 |
| Distributed | AL, CA, FL, IA, IL, MD, MI, MN, MO, MS, NC, NE, NJ, NY, PA, TX, UT, VA, WA, WI |
| Quantity | 1,662 |
| Reason classes | device malfunction |
| Lot numbers | 018810, 018850, 018860, 018880, 018900, 030090, 030100, 030140, 030150, 030180, 049860, 049870, 049900, 096520, 108200, 108290, 129340, 129420, 129440, 129460, 129500, 129520, 129550, 129570, 129580, 129590, 129600, 129620, 142280, 142290, 142310, 348480, 397350, 397370, 397400, 397430, 397440, 407250, 407260, 407270 and 108 more |
| Model numbers | 14-451070, 14-451075, 14-451080, 14-451085, 14-451090, 14-451095, 14-451100, 14-451105, 14-451110, 14-451115, 14-451120, 14-451125, 14-451130 |
Product
THP Hip Plating System, Item Nos. 14-451070 14-451075 14-451080 14-451085 14-451090 14-451095 14-451100 14-451105 14-451110 14-451115 14-451120 14-451125 14-451130 . The product is a metallic bone fixation system.
Reason for recall
Potential increase in fracture rate was detected during internal testing, in addition to potential discoloration inside the inner pouch of the packaging.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1347-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.