Data Foundry

Medical device recalls 2016

SOMATOM Definition AS — Class II medical device recall Z-1350-2016

Terminated recall by Siemens Medical Solutions USA, Inc, reported 2016-04-20. Siemens became aware that the front cover on some Siemens CT system was not properly welded. It is possible fo

Recall numberZ-1350-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Medical Solutions USA, Inc, Malvern, PA, United States
Recall initiated2016-03-22
Classified2016-04-08
Reported by FDA2016-04-20
Terminated2017-08-07
Quantity1 ct
Reason classesdevice malfunction

Product

SOMATOM Definition AS, a family of CT systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission dta from either the same axial plane taken at different angles or spiral planes taken at different angles.

Reason for recall

Siemens became aware that the front cover on some Siemens CT system was not properly welded. It is possible for the welded-in threaded bolt to break out of the ground plate of the front cover if the cover is opened during service. This could not happen during clinical operation. Therefore, there is no potential risk for the user or patient and the CT System has no risk for a malfunction during clinical operation