Abbott ARCHITECT Estradiol Reagent Kit — Class II medical device recall Z-1353-2019
Terminated recall by Abbott Ireland Diagnostics Division, reported 2019-05-22, distributed nationwide. Patient results may be falsely elevated. This patient impact only applies to patients currently being treated
| Recall number | Z-1353-2019 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Abbott Ireland Diagnostics Division, Co. Longford, Ireland |
| Recall initiated | 2019-02-05 |
| Classified | 2019-05-16 |
| Reported by FDA | 2019-05-22 |
| Terminated | 2020-08-21 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AR, AT, AU, BE, BR, CA, CH, CL, CN, CR, CW, CY, CZ, DE, DZ, EE, EG, ES, FI, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IR, IT, JM, JO, JP, KE, KR, KW, LB, LT, LU, LV, MA, MX, MY, NG, NL, NO, NZ, OM, PA, PE, PH, PK, PL, PT, QA, RO, RS, RU, SA, SE, SG, SI, SK, SV, TH, TN, TR, TW, VN, ZA |
| Quantity | 2,724 |
Product
Abbott ARCHITECT Estradiol Reagent Kit - Product Usage: The ARCHITECT Estradiol assay is a Chemiluminescent Microparticle Immunoassay (CMIA) for the quantitative determination of Estradiol in human serum and plasma.
Reason for recall
Patient results may be falsely elevated. This patient impact only applies to patients currently being treated with or recently treated with the drug Mifepristone.