Data Foundry

Medical device recalls 2026

da Vinci 5 Surgeon Console Viewer Display — Class II medical device recall Z-1355-2026

Ongoing recall by Intuitive Surgical, Inc., reported 2026-02-18, distributed nationwide. Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of

Recall numberZ-1355-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmIntuitive Surgical, Inc., Sunnyvale, CA, United States
Recall initiated2025-12-30
Classified2026-02-12
Reported by FDA2026-02-18
DistributedNationwide (US)
Quantity47
Reason classessoftware
UPC / GTIN / UDI-DI00886874119747
Serial numbersSQ0326, SQ0336, SQ0341, SQ0345, SQ0346, SQ0356, SQ0367, SQ0368, SQ0375, SQ0377, SQ0379, SQ0380, SQ0386, SQ0387, SQ0389, SQ0394, SQ0395, SQ0398, SQ0401, SQ0404, SQ0412, SQ0413, SQ0415, SQ0419, SQ0420 and 22 more
Model numbers380730-45

Product

da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45

Reason for recall

Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field (depth perception) but 2D maintained, or loss of both displays-complete visualization loss, without fault, instruments remain active, could lead to instrument contact with tissue, resulting in injury/bleeding; may require alternate surgical modality.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1355-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.