Data Foundry

Medical device recalls 2015

GAP Nail 4 — Class II medical device recall Z-1362-2015

Terminated recall by Pega Medical Inc., reported 2015-04-08, distributed nationwide. Failures of the 4.8mm of the GAP Endo-Exo Medullary System have been identified at the level of the distal cor

Recall numberZ-1362-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPega Medical Inc., Laval, N/A, Canada
Recall initiated2015-02-20
Classified2015-04-01
Reported by FDA2015-04-08
Terminated2015-04-30
DistributedNationwide (US)
Quantity7 units
Reason classesdevice malfunction
Lot numbers110913-001, 110913-002, 110913-004, 110913-005
Model numbersGAP-N48-16

Product

GAP Nail 4.8mm (diameter size) of the GAP ENDO-EXO MEDULLARY SYSTEM. Catalog number GAP-N48-16; GAP-N48-18; GAP-N48-26; GAP-N48-28. Product Usage: The GAP Endo-Exo Medullary System is used for the treatment of fractures or correction of deformities in the femur, tibia and humerus of pediatric patients, who have reached skeletal maturity. The GAP Nail is an intramedullary nail. It is the main component of the GAP Endo-Exo Medullary System. The rest of the components of the system are screws, plates and caps that are assembled to the nail during the surgical procedure. The GAP Endo-Exo Medullary System is indicated as a temporary implant to assure alignment, stabilization and fixation of: long bones that have been surgically prepared (osteotomy) for correction of deformities or fractures caused by trauma or disease. The GAP System is used for pediatric patients (child and adolescent) ages 2 to 21. It can be used to correct the following conditions: -Diaphyseal fracture of the femur, tibia and humerus -Fractures of the femoral neck -Subtrochanteric, intertrochanteric and combination fractures -Correction of deformities (OI, Coxa vara, Coxa valga) -Nonunions and malunions

Reason for recall

Failures of the 4.8mm of the GAP Endo-Exo Medullary System have been identified at the level of the distal cortical screw holes. Failures mainly occurred in patients with tibial pseudoarthrosis (CPT) or Osteogenesis imperfecta (OI). Higher failure rates have been detected in tibias with distal fractures that are close to the distal locking holes.