Teslatome Bipolar Sphincterotomes REF: TESLA-B2535-260 Model Number:… — Class II medical device recall Z-1362-2026

Ongoing recall by Wilson-Cook Medical Inc., reported 2026-02-25, distributed in CA, CO, IN, NC. Due to product that may contain a void in the material that separates the wire guide and cutting wire lumens,

Recall numberZ-1362-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmWilson-Cook Medical Inc., Winston Salem, NC, United States
Recall initiated2025-12-18
Classified2026-02-13
Reported by FDA2026-02-25
DistributedCA, CO, IN, NC
Quantity153
UPC / GTIN / UDI-DI00827002607842, 00827002607859
Model numbersG60784, G60785, TESLA-B2535-260, TESLA-B2535-450

Product

Teslatome Bipolar Sphincterotomes REF: TESLA-B2535-260 Model Number: G60784 UDI-DI code: 00827002607842 REF: TESLA-B2535-450 Model Number: G60785 UDI-DI code: 00827002607859 These devices are used for cannulation of the ductal system and for sphincterotomy. If preloaded, the device also aids in bridging difficult strictures during ERCP. It is indicated for sphincterotome-aided, wire-guided selective cannulation of the biliary ducts.

Reason for recall

Due to product that may contain a void in the material that separates the wire guide and cutting wire lumens, which may cause an inappropriate path of current during cutting.