Cobas¿ infinity central lab IT solution Material Number: 07154003001… — Class II medical device recall Z-1368-2019
Terminated recall by Roche Diagnostics Corporation, reported 2019-05-29, distributed in AL, AZ, CA, GA, IA, MO, NJ, NY…. Using the following versions of cobas infinity software (2.0 thorough 2.5), there is a potential where the res
| Recall number | Z-1368-2019 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Roche Diagnostics Corporation, Indianapolis, IN, United States |
| Recall initiated | 2019-02-04 |
| Classified | 2019-05-17 |
| Reported by FDA | 2019-05-29 |
| Terminated | 2020-09-16 |
| Distributed | AL, AZ, CA, GA, IA, MO, NJ, NY, SC, TN, TX |
| Quantity | 15 units |
| Reason classes | software |
| Serial numbers | 81191, 81371, 81385, 81465, 81469, 81470, 81481, 81485, 81527, 81528, 81618, 81628, 81652, 81655, 81662 |
Product
Cobas¿ infinity central lab IT solution Material Number: 07154003001 Calculator/Data Processing Module, For Clinical Use
Reason for recall
Using the following versions of cobas infinity software (2.0 thorough 2.5), there is a potential where the results from the first sample subsequent to the bottle change, (when using the Standby bottle function), does not display a quality control (QC) alarm when one should be present. If the QC is out of range on the stand-by bottle and it is not removed from the system, there is a potential that when the status of the Standby bottle changes from the Standby bottle to the Current bottle status, QC alarms will not be displayed next to the sample result.