Data Foundry

Medical device recalls 2018

Hamilton-C3 Ventilator Unit Intended to provide ventilator support to… — Class II medical device recall Z-1372-2018

Ongoing recall by Hamilton Medical AG, reported 2018-04-18, distributed in AZ, CA, CO, FL, GA, HI, IL, IN…. After two and a half years, the battery fuel gauge may indicate a higher battery charge than what is available

Recall numberZ-1372-2018
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmHamilton Medical AG, Bonaduz, Switzerland
Recall initiated2018-03-12
Classified2018-04-12
Reported by FDA2018-04-18
DistributedAZ, CA, CO, FL, GA, HI, IL, IN, KY, MA, MI, MN, MO, MT, NC, ND, NE, NV, OH, OK, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA
Countries outside the USAE, AR, AT, AU, BD, BG, BO, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DZ, EC, EG, ES, FI, FR, GB, GR, GT, HN, HR, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KR, KW, LB, LK, LT, LV, MA, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, RO, RS, RU, SA, SE, SI, SK, SV, TH, TN, TR, TT, TW, UA, UY, VE, VN, ZA
Quantity739 units
Serial numbersHAMILTON-C3

Product

Hamilton-C3 Ventilator Unit Intended to provide ventilator support to adults and pediatrics and optionally infants and neonates in the hospital.

Reason for recall

After two and a half years, the battery fuel gauge may indicate a higher battery charge than what is available.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1372-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.