VITROS Performance Verifier II — Class II medical device recall Z-1373-2024
Ongoing recall by Ortho-Clinical Diagnostics, reported 2024-04-03, distributed nationwide. The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. T
| Recall number | Z-1373-2024 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ortho-Clinical Diagnostics, Rochester, NY, United States |
| Recall initiated | 2024-02-27 |
| Classified | 2024-03-22 |
| Reported by FDA | 2024-04-03 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, BE, BM, BR, CA, CL, CN, CO, DE, DK, ES, FR, GB, IN, IT, JP, MX, NL, NO, PT, RU, SE, SG |
| Quantity | 11,224 units |
| UPC / GTIN / UDI-DI | 10758750004577 |
| Lot numbers | R1176 |
Product
VITROS Performance Verifier II
Reason for recall
The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause erroneous Quality Control (QC) results that could lead to testing delays and erroneous patient results. This issue caused positive or negative bias in the quantification of sodium (Na+) and total carbon dioxide (ECO2). The incorrect diluent into the Performance Verifier II exhibited a potential negative bias of up to 6 mmol/L for ECO2 and Na+ testing.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1373-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.