Obsidio Conformable Embolic — Class I medical device recall Z-1374-2024

Ongoing recall by Boston Scientific Corporation, reported 2024-04-03. An investigation determined that delivery of the Obsidio embolic using the aliquot technique for lower gastroi

Recall numberZ-1374-2024
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBoston Scientific Corporation, Maple Grove, MN, United States
Recall initiated2024-02-20
Classified2024-03-27
Reported by FDA2024-04-03
Quantity985 units
UPC / GTIN / UDI-DI00191506039332, 00191506043124

Product

Obsidio Conformable Embolic, Material Numbers (UPN): a) M0013972001010, b) M0013972101010

Reason for recall

An investigation determined that delivery of the Obsidio embolic using the aliquot technique for lower gastrointestinal bleeding embolization poses a high risk of bowel ischemia. The most serious and the most common adverse health consequence, reasonably foreseeable to occur, is the need to perform major surgery such as bowel resection and/or diverting colostomy. Therefore, Boston Scientific does not recommend that the aliquot technique be used to deliver the Obsidio device for lower GI bleed embolization procedures.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1374-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.