Data Foundry

Medical device recalls 2018

MINC Benchtop Incubator — Class II medical device recall Z-1378-2018

Terminated recall by Cook Inc., reported 2018-04-25, distributed nationwide. The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for pro

Recall numberZ-1378-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmCook Inc., Bloomington, IN, United States
Recall initiated2018-01-12
Classified2018-04-17
Reported by FDA2018-04-25
Terminated2020-05-08
DistributedNationwide (US)
Countries outside the USAU
Quantity4,451
Model numbersG20079

Product

MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000) G20079 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.

Reason for recall

The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1378-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.