Data Foundry

Medical device recalls 2021

Trial Cutting Guide for Triathlon Total Knee System — Class II medical device recall Z-1379-2021

Terminated recall by Howmedica Osteonics Corp., reported 2021-04-14, distributed in CA, GA, NC, NJ, NV, OK, TX. The posterior condyle of the Specialty Triathlon TS Trial Cutting Guide(s), sizes 1 through 8, (Left and Right

Recall numberZ-1379-2021
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmHowmedica Osteonics Corp., Mahwah, NJ, United States
Recall initiated2021-03-05
Classified2021-04-08
Reported by FDA2021-04-14
Terminated2024-05-10
DistributedCA, GA, NC, NJ, NV, OK, TX
Countries outside the USCA
Quantity113 components
Reason classesdevice malfunction
UPC / GTIN / UDI-DI07613327327533, 07613327327595, 07613327327748, 07613327328288, 07613327328424, 07613327328530, 07613327328592, 07613327328639, 07613327328691, 07613327328707, 07613327328868, 07613327328899, 07613327329063, 07613327329087, 07613327329100, 07613327329193, 07613327329209
Model numbersI-K3834CG00

Product

Trial Cutting Guide for Triathlon Total Knee System, Triathlon PKR System, Restoris MCK Knee System, prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer. Catalog Number: I-K3834CG00 Specialty Triathlon TS Trial Cutting Guide Set I-K3834CG1L Specialty Triathlon TS Trial Cutting Guide Size 1 Left I-K3834CG1R Specialty Triathlon TS Trial Cutting Guide Size 1 Right I-K3834CG2L Specialty Triathlon TS Trial Cutting Guide Size 2 Left I-K3834CG2R Specialty Triathlon TS Trial Cutting Guide Size 2 Right I-K3834CG3L Specialty Triathlon TS Trial Cutting Guide Size 3 Left I-K3834CG3R Specialty Triathlon TS Trial Cutting Guide Size 3 Right I-K3834CG4L Specialty Triathlon TS Trial Cutting Guide Size 4 Left I-K3834CG4R Specialty Triathlon TS Trial Cutting Guide Size 4 Right I-K3834CG5L Specialty Triathlon TS Trial Cutting Guide Size 5 Left I-K3834CG5R Specialty Triathlon TS Trial Cutting Guide Size 5 Right I-K3834CG6L Specialty Triathlon TS Trial Cutting Guide Size 6 Left I-K3834CG6R Specialty Triathlon TS Trial Cutting Guide Size 6 Right I-K3834CG7L Specialty Triathlon TS Trial Cutting Guide Size 7 Left I-K3834CG7R Specialty Triathlon TS Trial Cutting Guide Size 7 Right I-K3834CG8L Specialty Triathlon TS Trial Cutting Guide Size 8 Left I-K3834CG8R Specialty Triathlon TS Trial Cutting Guide Size 8 Right

Reason for recall

The posterior condyle of the Specialty Triathlon TS Trial Cutting Guide(s), sizes 1 through 8, (Left and Right) has the potential to fracture at the augment cut slot. This can occur in the Trial Cutting Guide packaging during transport or during impaction of the instrument in surgery.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1379-2021. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.