Data Foundry

Medical device recalls 2018

Getinge AGS — Class II medical device recall Z-1380-2018

Terminated recall by Getinge Disinfection Ab, reported 2018-04-25, distributed nationwide. There is potential for a cart with disinfected goods to unload without the trolley being docked. Disinfected g

Recall numberZ-1380-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmGetinge Disinfection Ab, Vaxjo, Sweden
Recall initiated2018-03-02
Classified2018-04-16
Reported by FDA2018-04-25
Terminated2022-05-13
DistributedNationwide (US)
Countries outside the USAU, BE, BR, CA, CN, CO, CZ, DE, DK, ES, FI, FR, GB, HK, IT, JP, KR, KW, MY, NL, NO, SA, SE, SG, SI, TH
Quantity170 products
UPC / GTIN / UDI-DI07340153700109, 07340153700116
Serial numbersWAA056686, WAA057547, WAA058540, WAA058542, WAA058543, WAA058545, WAA058549, WAA058550, WAA058551, WAA058552, WAA058553, WAA058554, WAA059053, WAA059054, WAA059058, WAA059059, WAA059509, WAA059510, WAA059630, WAA059785, WAA060117, WAA060690, WAA060871, WAA060886, WAA060887 and 135 more

Product

Getinge AGS, FSLC, FSUC, RC accessory to Washer- Disinfector 88 Series UDI Info: 07340153700109, 07340153700116 Getinge automated loading/unloading systems are intended to be used as an automatic conveyor/transport system for unloading carriers (wash carts) from Getinge 86/88 series Washer-disinfectors. The automated loading/unloading systems in no way affect the cleaning/disinfection performance of the washer-disinfector. The automated loading/unloading equipment is considered an accessory to the medical washer-disinfector.

Reason for recall

There is potential for a cart with disinfected goods to unload without the trolley being docked. Disinfected goods could end up falling to the floor causing risk to nearby users and damage to goods and equipment.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1380-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.