Data Foundry

Medical device recalls 2023

DigitalDiagnost C50 1 — Class II medical device recall Z-1383-2023

Ongoing recall by Philips North America Llc, reported 2023-04-19, distributed nationwide. If the PM for the telescopic carriage is not performed annually as defined in the Instructions for Use (IFU),

Recall numberZ-1383-2023
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPhilips North America Llc, Cambridge, MA, United States
Recall initiated2023-03-09
Classified2023-04-12
Reported by FDA2023-04-19
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BG, CH, CN, CO, CY, CZ, DE, EG, ES, FR, GR, HU, ID, IT, JO, KE, LK, LT, LV, MT, MX, NL, NZ, PH, PL, PT, RS, SA, SI, TH, VN, ZA
UPC / GTIN / UDI-DI00884838095199
Serial numbers200001, 200002, 200005, 200006, 200007, 200009, 200010, 200011, 200012, 200013, 200014, 200015, 200017, 200018, 200019, 200022, 200023, 200024, 200026, 200027, 200028, 200029, 200030, 200031, 200032 and 268 more
Model numbers712204

Product

DigitalDiagnost C50 1.1-intended for use in generating radiographic images of human anatomy Model Number : 712204

Reason for recall

If the PM for the telescopic carriage is not performed annually as defined in the Instructions for Use (IFU), the tension force of the spring balancer inside of the telescopic carriage will unavoidably decay through normal wear and tear; as a result there is a potential that the Ceiling Suspension of the DigitalDiagnost C50 system may descend inadvertently if power to the system is interrupted, and may result in injuries to patients and users

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1383-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.