Data Foundry

Medical device recalls 2019

UniCel DxH 900 Coulter Cellular Analysis System — Class I medical device recall Z-1384-2019

Terminated recall by Beckman Coulter Inc., reported 2019-05-29, distributed nationwide. Complaints received of sporadic erroneously elevated platelet results without flags or system messages. Thromb

Recall numberZ-1384-2019
Product typeMedical device
ClassificationClass I
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBeckman Coulter Inc., Brea, CA, United States
Recall initiated2018-07-30
Classified2019-05-23
Reported by FDA2019-05-29
Terminated2022-03-31
DistributedNationwide (US)
Countries outside the USAE, AT, AU, BE, CA, CH, CN, DE, EG, ES, FR, GB, GR, HR, HU, IL, IN, IT, JP, KR, KW, LT, LV, MA, MX, NL, NZ, PT, RO, RS, RU, SA, SI, TT, VN
Quantity341
Model numbersC11478

Product

UniCel DxH 900 Coulter Cellular Analysis System, REF number C11478 Product Usage: Note: The DxH 900 hematology analyzer is CLIA categorized under document CR180294 and is currently marketed as a member of the instrument family of the DxH 800 cleared under K140911, per FDA s guidance, Guidance for Industry and FDA Staff, Replacement Reagent and Instrument Family Policy . DxH 800 intended use states: The UniCel DxH 800 Analyzer is a quantitative multi-parameter, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The UniCel¿ DxH 800 Analyzer identifies and enumerates the parameters indicated below on the following sample types: Whole Blood (Venous and Capillary) o WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV, NE%, NE#, LY%, LY#, MO%, MO#, EO%, EO#, BA%, BA#, NRBC%, NRBC#, RET%, RET#, MRV, IRF Pre-Diluted Whole Blood (Venous and Capillary) o WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV Body Fluids (cerebrospinal, serous and synovial) o TNC and RBC

Reason for recall

Complaints received of sporadic erroneously elevated platelet results without flags or system messages. Thrombocytopenia may go unrecognized.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1384-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.