VERICIS — Class II medical device recall Z-1398-2025
Ongoing recall by Merge Healthcare, Inc., reported 2025-03-26. When following a specific less typical workflow in which the user remeasures using the ultrasound device or ma
| Recall number | Z-1398-2025 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Merge Healthcare, Inc., Hartland, WI, United States |
| Recall initiated | 2025-02-19 |
| Classified | 2025-03-20 |
| Reported by FDA | 2025-03-26 |
| Quantity | 2 |
| Model numbers | 88-00054-00, 88-00061-00, 88-00062-00, 88-00063-00, 88-00064-00 |
Product
VERICIS, Merge Cardio, Model/Catalog Number: Version 9.x, Software Version: 9.0.6 and 9.0.8; VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.
Reason for recall
When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1398-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.