Data Foundry

Medical device recalls 2015

ADVIA Chemistry XPT System — Class II medical device recall Z-1399-2015

Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2015-04-15, distributed in AZ, CA, WA. Siemens Healthcare Diagnostics has confirmed an issue with ADVIA Chemistry XPT software version 1.0.2. The sys

Recall numberZ-1399-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc., Tarrytown, NY, United States
Recall initiated2015-02-11
Classified2015-04-07
Reported by FDA2015-04-15
Terminated2017-03-02
DistributedAZ, CA, WA
Countries outside the USDE, ES, GB, IT
Reason classessoftware
Lot numbersCA1275000050005, CA1275000060006, CA1275000070007, CA1275000080008, CA1275000090009, CA1275000100010, CA1275000120012, CA1275000130013, CA1275000170017, CA1275000200020, CA1275000220022, CA1275000240024, CA1275000260026, CA1275000270027, CA1275000280028, CA1275000410041

Product

ADVIA Chemistry XPT System, Siemens Material Number 10723034, Software Version 1.0.2, Siemens Material Number 11219493, IVD. The ADVIA XPT Chemistry System is an automated, clinical chemistry analyzer that runs tests on serum, plasma, urine, or cerebral spinal fluid in random access and batch modes at a throughput rate of both 1800 photometric tests per hour and 600 electrolyte (ISE) tests per hour.

Reason for recall

Siemens Healthcare Diagnostics has confirmed an issue with ADVIA Chemistry XPT software version 1.0.2. The system may cause samples to remain in an Inprocess state. Test results on a sample that is held Inprocess will not transmit to the LIS. Manual intervention is necessary to complete the processing of the samples that are held Inprocess.