Data Foundry

Medical device recalls 2025

VERICIS — Class II medical device recall Z-1401-2025

Ongoing recall by Merge Healthcare, Inc., reported 2025-03-26. When following a specific less typical workflow in which the user remeasures using the ultrasound device or ma

Recall numberZ-1401-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMerge Healthcare, Inc., Hartland, WI, United States
Recall initiated2025-02-19
Classified2025-03-20
Reported by FDA2025-03-26
Quantity37
UPC / GTIN / UDI-DI00842000100768

Product

VERICIS, Merge Cardio, Model/Catalog Number: Version 12.X, Software Version: 12.0.1, 12.0.2; The VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.

Reason for recall

When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1401-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.