Data Foundry

Medical device recalls 2018

Brilliance Big Bore — Class II medical device recall Z-1411-2018

Terminated recall by Philips Medical Systems (Cleveland) Inc, reported 2018-04-25, distributed nationwide. Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient loa

Recall numberZ-1411-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips Medical Systems (Cleveland) Inc, Cleveland, OH, United States
Recall initiated2018-02-08
Classified2018-04-17
Reported by FDA2018-04-25
Terminated2020-04-09
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BD, BE, BG, BO, BR, CA, CH, CL, CN, CO, CZ, DE, DK, DO, DZ, EE, EG, ES, FI, FR, GB, GH, GR, GY, HK, HR, HU, ID, IE, IL, IN, IR, IT, JO, JP, KE, KR, LB, LT, MX, MY, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, QA, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, VE, VN, ZA
Quantity119
Serial numbers750003, 750011, 75030, 75033, 75043, 75106, 75110, 75117, 75122, 75123, 75124, 75125, 760001, 760002, 760003, 760004, 760005, 760006, 760007, 760008, 760009, 76001, 760010, 760011, 760012 and 90 more
Model numbers728244

Product

Brilliance Big Bore (Radiology) - Model no. 728244, Computed Tomography X-Ray system Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories.

Reason for recall

Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1411-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.