Natus neoBLUE blanket LED Phototherapy system which includes the… — Class II medical device recall Z-1412-2015
Terminated recall by Natus Medical Incorporated, reported 2015-04-22. neoBLUE blanket LED Phototherapy System is recalled because Natus has received feedback from some customers ab
| Recall number | Z-1412-2015 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Natus Medical Incorporated, Seattle, WA, United States |
| Recall initiated | 2015-03-16 |
| Classified | 2015-04-10 |
| Reported by FDA | 2015-04-22 |
| Terminated | 2016-12-20 |
| Model numbers | 006224 |
Product
Natus neoBLUE blanket LED Phototherapy system which includes the following items: 006224 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad 006254 neoBLUE blanket LED Light Box 006895 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad 007299 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad EUR 007300 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad EUR 007296 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad UK 007298 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad UK 007301 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad AUS 007302 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad AUS
Reason for recall
neoBLUE blanket LED Phototherapy System is recalled because Natus has received feedback from some customers about early failure of the neoBLUE blanket system fiberoptic pads. These failures involve discoloration and melting of the fiberoptic bundle at the connector that is inserted in the neoBLUE blanket light box.