Data Foundry

Medical device recalls 2023

BD Kiestra InoqulA+ with BD Kiestra" InoqulA+" and BeA Software… — Class II medical device recall Z-1415-2023

Ongoing recall by Becton Dickinson & Co., reported 2023-04-26, distributed nationwide. Upon installation of BD Kiestra" InoqulA+" and BeA version 5.1 and 5.1.1 as part of the Icefall A platform, pl

Recall numberZ-1415-2023
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBecton Dickinson & Co., Sparks, MD, United States
Recall initiated2023-03-22
Classified2023-04-20
Reported by FDA2023-04-26
DistributedNationwide (US)
Countries outside the USNL, SA
Quantity7 instruments
Serial numbers19033ST002, 20130276, INO-000156, INO-000248, INO-000249, INO-000252, INO-000254
Model numbers447213

Product

BD Kiestra InoqulA+ with BD Kiestra" InoqulA+" and BeA Software version 5.1 and 5.1.1 as part of the Icefall A platform. An in vitro diagnostic medical device intended to automate specimen processing according to user-defined procedures and protocols. Catalog (Ref) No. : 447213

Reason for recall

Upon installation of BD Kiestra" InoqulA+" and BeA version 5.1 and 5.1.1 as part of the Icefall A platform, plate information was not visible in Synapsys" after processing has occurred. This could possibly lead to a delayed diagnosis and/or treatment or need for specimen recollection streaking plated media, and inoculating tubes and slides. In Semi-Automated (SA) mode, plates are automatically selected, barcoded, streaked in a pre-configured pattern while the user manually inoculates plates, tubes, and slides. An optional biosafety cabinet on the SA module provides personnel, product, and environmental protection. The InoqulA+" solution is indicated for use in the clinical laboratory.