Data Foundry

Medical device recalls 2018

Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride… — Class II medical device recall Z-1420-2018

Ongoing recall by Autochair, reported 2018-04-25. Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the lo

Recall numberZ-1420-2018
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmAutochair, Alfreton, United Kingdom
Recall initiated2018-03-07
Classified2018-04-18
Reported by FDA2018-04-25
Model numbers9934105

Product

Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B

Reason for recall

Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards while lifting a person off a bed and the individual was dropped.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1420-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.