Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride… — Class II medical device recall Z-1420-2018
Ongoing recall by Autochair, reported 2018-04-25. Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the lo
| Recall number | Z-1420-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Autochair, Alfreton, United Kingdom |
| Recall initiated | 2018-03-07 |
| Classified | 2018-04-18 |
| Reported by FDA | 2018-04-25 |
| Model numbers | 9934105 |
Product
Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B
Reason for recall
Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards while lifting a person off a bed and the individual was dropped.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1420-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.