LifeGlobal Fast Freeze Thawing Kit — Class II medical device recall Z-1421-2022

Ongoing recall by CooperSurgical, Inc., reported 2022-07-27, distributed in AL, AZ, CA, CT, FL, GA, ID, IL…. There was translation error in the Spanish version of the IFU where the Spanish version instructs the user to

Recall numberZ-1421-2022
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmCooperSurgical, Inc., Trumbull, CT, United States
Recall initiated2022-06-13
Classified2022-07-20
Reported by FDA2022-07-27
DistributedAL, AZ, CA, CT, FL, GA, ID, IL, IN, KS, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TX, UT, VT, WA, WI
Countries outside the USBE, CA, ES, HR, NL, PT
Quantity403 units
Reason classestemperature control
UPC / GTIN / UDI-DI00815965020914
Lot numbers21335060, GFT5-200218U, GFT5-201026U
Model numbersGFT5-055

Product

LifeGlobal Fast Freeze Thawing Kit, Model No. GFT5-055

Reason for recall

There was translation error in the Spanish version of the IFU where the Spanish version instructs the user to leave the blastocyte to air thaw for 5 minutes rather than the correct 5 seconds as indicated in the English version.