Data Foundry

Medical device recalls 2026

Medline Convenience Kits: 1) SPINAL CDS — Class II medical device recall Z-1421-2026

Ongoing recall by Medline Industries, LP, reported 2026-02-25, distributed nationwide. Medline has identified issues related to calibration of the equipment used to sterilize and package the device

Recall numberZ-1421-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMedline Industries, LP, Northfield, IL, United States
Recall initiated2026-01-07
Classified2026-02-17
Reported by FDA2026-02-25
DistributedNationwide (US)
Quantity117,907 units
Reason classessterility
UPC / GTIN / UDI-DI10080196979752, 10193489201055, 10193489201901, 10193489207439, 10193489210934, 10193489216455, 10193489219807, 10193489220261, 10193489220698, 10193489222623, 10193489222852, 10193489226768, 10193489234664, 10193489237528, 10193489242669, 10193489245370, 10193489245424, 10193489245639, 10193489245820, 10193489246858, 10193489249705, 10193489270976, 10193489271607, 10193489272017, 10193489279146 and 475 more
Lot numbers100, 1000, 1001, 1002, 1003, 1004, 1005, 1006, 1007, 1008, 1009, 101, 1010, 1011, 1012, 1013, 1014, 1015, 1016, 1017, 1018, 1019, 102, 1020, 1021, 1022, 1023, 1024, 1025, 1026, 1027, 1028, 1029, 103, 1030, 1031, 1032, 1033, 1034, 1035 and 460 more
Model numbersCDS940087AG, CDS940087AI, CDS980832Q, CDS980832R, CDS980832S, CDS980833Q, CDS980833R, CDS980865V, CDS980865W, CDS980865X, CDS980865Y, CDS981454K, CDS981454L, CDS981502K, CDS981505O, CDS981505P, CDS981505Q, CDS981505R, CDS981858F, CDS982035W, CDS982281O, CDS982281P, CDS982281S, CDS983036K, CDS983042K and 39 more

Product

Medline Convenience Kits: 1) SPINAL CDS, Model Number: CDS940087AG; 2) SPINAL CDS, Model Number: CDS940087AI; 3) TOTAL HIP CDS-LF, Model Number: CDS980832Q; 4) TOTAL HIP CDS-LF, Model Number: CDS980832R; 5) TOTAL HIP CDS-LF, Model Number: CDS980832S; 6) TOTAL KNEE CDS-LF, Model Number: CDS980833Q; 7) TOTAL KNEE CDS-LF, Model Number: CDS980833R; 8) TOTAL HIP PROCEDURE, Model Number: CDS980865V; 9) TOTAL HIP PROCEDURE, Model Number: CDS980865W; 10) TOTAL HIP PROCEDURE, Model Number: CDS980865X; 11) TOTAL HIP PROCEDURE, Model Number: CDS980865Y; 12) TOTAL KNEE CDS, Model Number: CDS981454K; 13) TOTAL KNEE CDS, Model Number: CDS981454L; 14) KNEE ARTHROSCOPY CDS, Model Number: CDS981502K; 15) TOTAL KNEE, Model Number: CDS981505O; 16) TOTAL KNEE, Model Number: CDS981505P; 17) TOTAL KNEE, Model Number: CDS981505Q; 18) TOTAL KNEE, Model Number: CDS981505R; 19) OR LAMINECTOMY CDS, Model Number: CDS981858F; 20) LAMI FUSION, Model Number: CDS982035W; 21) ARMC TOTAL KNEE CDS, Model Number: CDS982281O; 22) ARMC TOTAL KNEE CDS, Model Number: CDS982281P; 23) ARMC TOTAL KNEE CDS, Model Number: CDS982281S; 24) TOTAL HIP CDS, Model Number: CDS983036K; 25) TOTAL KNEE CDS, Model Number: CDS983042K; 26) LOWER EXTREMITY CDS, Model Number: CDS983049K; 27) ZALE SPINE CDS, Model Number: CDS983182G; 28) TOTAL KNEE CDS, Model Number: CDS983297M; 29) SITTING CERVICAL CDS, Model Number: CDS983480G; 30) SHOULDER CDS, Model Number: CDS983581K; 31) SHOULDER ARTHROSCOPY CDS, Model Number: CDS983838G; 32) TOTAL HIP CDS, Model Number: CDS983840I; 33) TOTAL JOINT CDS-LF, Model Number: CDS983863D; 34) SHOULDER ARTHROSCOPY CDS, Model Number: CDS983879C; 35) TOTAL JOINT, Model Number: CDS983903D; 36) GB TOTAL KNEE, Model Number: CDS984147K; 37) GB LOWER EXTREMITY, Model Number: CDS984156L; 38) LOWER EXTREMTIY, Model Number: CDS984254K; 39) LOWER EXTREMTIY, Model Number: CDS984254L; 40) LOWER EXTREMTIY, Model Number: CDS984254M; 41) LAMI MICRODISC-LF, Model Number: CDS984527I; 42) EXTREMITY CDS-LF, Model Number: CDS984607G; 43) EXTREMITY CDS-LF, Model Number: CDS984607I; 44) HANA/FRACTURE TABLE CDS-LF, Model Number: CDS984610C; 45) TOTAL JOINT CDS, Model Number: CDS985202B; 46) TOTAL JOINT CDS, Model Number: CDS985202C; 47) TRAUMA-LF, Model Number: CDS985279K; 48) ARTHROSCOPY CDS, Model Number: CDS985316A; 49) EXTREMITY CDS, Model Number: CDS985317C; 50) EXTREMITY CDS, Model Number: CDS985317D; 51) EXTREMITY CDS, Model Number: CDS985317F; 52) EXTREMITY CDS, Model Number: CDS985317G; 53) KNEE ARTHROSCOPY CDS, Model Number: CDS985425J; 54) TOTAL HIP PACK (CDSOPM834)-LF, Model Number: CDSOPM834W; 55) JOINT ASPIRATION KIT, Model Number: DYKM1161; 56) TOTAL KNEE SUPPLEMENT PACK NS, Model Number: DYKM2106; 57) KIT ORTHRO TOTAL HIP DR. NANDI, Model Number: DYKMBNDL137; 58) KIT NEUR FUS LUM POST CRANDALL, Model Number: DYKMBNDL138A; 59) KIT NEURO CRANIOTOMYANEURYSM R, Model Number: DYKMBNDL141; 60) KIT NEURO CRANIOTOMYANEURYSM R, Model Number: DYKMBNDL141A; 61) KIT NEURO CRANIOTOMYANEURYSM R, Model Number: DYKMBNDL141B; 62) KIT NEURO FUS LUM POST SANSUR, Model Number: DYKMBNDL142C; 63) KIT NEURO CRANIOTOMY ACOUSTIC, Model Number: DYKMBNDL143; 64) KIT NEURO CRANIOTOMY FOR TUMOR, Model Number: DYKMBNDL144; 65

Reason for recall

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1421-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.