Data Foundry

Medical device recalls 2020

Triglycerides Reagent — Class II medical device recall Z-1423-2020

Terminated recall by Beckman Coulter Inc, reported 2020-03-11, distributed in AK, AL, AR, AZ, CA, CO, CT, DE…. It has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen (paracetamol)

Recall numberZ-1423-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBeckman Coulter Inc, Carlsbad, CA, United States
Recall initiated2019-08-15
Classified2020-03-02
Reported by FDA2020-03-11
Terminated2024-02-28
DistributedAK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY
Countries outside the USAE, AU, BD, BE, BG, CA, CH, CN, CW, CZ, DE, EG, ES, FI, FR, GB, GH, GR, GT, GY, HK, IE, IN, IT, JP, KR, KW, LB, LT, MX, MY, NL, OM, PA, PH, PL, PT, PY, QA, RO, RS, RU, SA, SG, TH, TN, TR, TW, VN, ZA
UPC / GTIN / UDI-DI15099590233228
Model numbers445850

Product

Triglycerides Reagent (TG 2 x 300), REF: 445850, For In Vitro Diagnostic Use, Rx Only, UDI: 15099590233228

Reason for recall

It has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen (paracetamol), may cause negative interference for assays if present in high quantities in serum due to acetaminophen overdoese. This could result in a delay in diagnosis and treatment.